Bettersizer 2600 - Wet or Dry Particle Size Analyzer

The Bettersizer 2600 from Bettersize Instruments can measure particle size using either a wet or dry method.

This versatile, powerful particle size analyzer is suitable for a wide variety of measurement applications thanks to its modular design and patented technologies. The Bettersizer 2600 allows users to easily and accurately characterize materials from 0.02 μm to 2600 μm.

Features and Benefits

  • Powerful laser diffraction technology
  • Wide particle size range, able to measure wet dispersion from 0.02 - 2600 μm and dry dispersion from 0.1 - 2600 μm
  • Five state-of-the-art wet and dry dispersion modules accommodate a diverse array of dispersing particle systems.
  • New, patented technology employing Fourier and Inversed Fourier optical systems enables the precise measurement of large and small particles across a broad range.
  • Robust optical bench featuring 92 detectors, covering an angular range of 0.016 ° - 165 °
  • A small volume dry dispersing module can accommodate small sample quantities - ideal for pharmaceutical samples or valuable materials in short supply.
  • Rapid, straightforward changeover between dispersing modules
  • Users can create and implement Standard Operation Procedures for new materials with ease.
  • Refractive index measurements offer a more accurate parameter for result calculations.
  • Easy-to-learn, user-friendly software

Specifications

Source: Bettersize Instruments Ltd.

Bettersizer 2600
Particle size distribution Suspension, emulsion, dry powders
General
Principle Laser diffraction technology
Analysis Mie scattering theory and Fraunhofer diffraction theory
Typical measurement time Less than 10 seconds
Measurement Performance
Measuring range 0.02 - 2600 μm (wet) *
0.1 - 2600 μm (dry) *
Accuracy error ≤ 0.5% *
Repeatability ≤ 0.5% *
Number of size classes 100 (user adjustable)
Feeding mode Automatic circulation or micro cuvette (wet)
Gas transportation (dry)
Special functions SOP settings, refractive index measurement, sample ratio calculation
Main Device
Optical system Combined Fourier and inverse Fourier & Tilted sample cell
Laser High-power optical fiber laser (10 mW / 635 nm)
Detector 92 detectors (forward, lateral and backward arrangements)
Measuring angle 0.016 - 165°
Wet dispersion module
Measurement method One-click operation (automated measurement, cleaning, saving, printing, etc.)
Dispersion medium Water or organic solvents (sample dependent)
Circulation speed 300 - 2500 r/min
Ultrasonication Dry burning prevention,50 W
Dry dispersion module
Dispersion medium Air/ Nitrogen/Noble gas
Air pressure 0.1 - 0.8 MPa (Air compressor dependent)
Software 
Conformity  21 CFR Part 11, ISO 13320, USP <429>
Report  Customizable reporting
System Parameters
Dimensions (L x W x H)  70.5 x 31.8 x 29.5 cm
Weight  23 kg
Supply voltage 100 - 240 V, 50/60 Hz
Computer Configuration (Recommended)
Computer interface  At least one high-speed USB 2.0 or USB 3.0 port required 
Operating system  Windows 7 (32-bit and 64-bit) or higher
Hardware specification  Intel Core i5 Processor, 4 GB RAM, 250 GB HD, Widescreen monitor

*Sample and sample preparation dependent

Bettersizer 2600 | Laser Diffraction Particle Size Analyzer (Wet & Dry)

Video Credit: Bettersize Instruments Ltd.

Other Equipment by this Supplier

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.